| Line Master (Covered Foot Pedal) | 
| #1 | 
  | 
| Problem | 
Ocusystem not responding to foot switch. | 
| Solution | 
- The Foot pedal cord is properly connected to the stand.
 
- The 18 inch interconnecting cable is connected from the back of the machine to the back of the stand.
 
- The pins in the Foot pedal plug are not broken.
 
- The Foot pedal cable is not torn or wires exposed.
 
- All indicator lights (right hand side of machine front panel) follow foot switch positions.
  
 | 
| #2 | 
  | 
| Problem | 
No Reflux. | 
| Solution | 
- Check that the plate in the hood of the foot pedal is in the proper position and activates the switch.
 
- Fix the plate by pushing it in the proper position.
 
- Make sure that there is fluid in the aspiration line, no air.
  
 | 
|   | 
| Open Style Foot Pedal (Advantage and Dual Linear) | 
| #1 | 
  | 
| Problem | 
Advantage not responding to foot switch. | 
| Solution | 
- The Foot pedal cord is properly connected to the stand.
 
- The 18 inch interconnecting cable is connected from the back of the machine to the back of the stand.
 
- The pins in the Foot pedal plug are not broken.
 
- Check the Foot pedal cable for broken or exposed wires.
 
- All indicator lights (right hand side of machine front panel) follow foot switch positions.
  
 | 
| #2 | 
  | 
| Problem | 
Wrong Memory or Function. | 
| Solution | 
- Make sure the surgeon has not press the function and memory switches on the foot pedal.
  
 |