| #1 | 
  | 
| Problem | 
No Phaco Power. | 
| Solution | 
- Check that the console is set for Phaco and the desired power level is set.
 
- Check that the Phaco plug is secure in the connector in the front panel.
 
- Check for broken wires near the plug and rear end of hand piece.
 
- Check that Phaco tip is tight.
 
- Check that cocoon is loose within the hand piece.
 
- Make sure that the Open Infusion button is not open during Phaco.
 
- Make that the Phaco power light is on when the surgeon goes in to the third position of the foot switch.
  
 | 
| #2 | 
  | 
| Problem | 
Clouding in Anterior Chamber. | 
| Solution | 
- Increase flow rate.
  
 | 
| #3 | 
  | 
| Problem | 
Incision Heating. | 
| Solution | 
- Make sure size of the incision is correct for phaco tip.
 
- Replace Cobra Sleeve, internal ribs may be worn.
 
- Before starting ultrasound make sure you have flow and tip is not clogged with viscoelastic.
 
- Make sure that the phaco tip is in the center of the incision.
  
 |