Line Master (Covered Foot Pedal) |
#1 |
|
Problem |
Ocusystem not responding to foot switch. |
Solution |
- The Foot pedal cord is properly connected to the stand.
- The 18 inch interconnecting cable is connected from the back of the machine to the back of the stand.
- The pins in the Foot pedal plug are not broken.
- The Foot pedal cable is not torn or wires exposed.
- All indicator lights (right hand side of machine front panel) follow foot switch positions.
|
#2 |
|
Problem |
No Reflux. |
Solution |
- Check that the plate in the hood of the foot pedal is in the proper position and activates the switch.
- Fix the plate by pushing it in the proper position.
- Make sure that there is fluid in the aspiration line, no air.
|
|
Open Style Foot Pedal (Advantage and Dual Linear) |
#1 |
|
Problem |
Advantage not responding to foot switch. |
Solution |
- The Foot pedal cord is properly connected to the stand.
- The 18 inch interconnecting cable is connected from the back of the machine to the back of the stand.
- The pins in the Foot pedal plug are not broken.
- Check the Foot pedal cable for broken or exposed wires.
- All indicator lights (right hand side of machine front panel) follow foot switch positions.
|
#2 |
|
Problem |
Wrong Memory or Function. |
Solution |
- Make sure the surgeon has not press the function and memory switches on the foot pedal.
|