Care & Maintenance: Pressure Transducer

At the end of each surgical day disinfect the Pressure Transducer:
  1. Connect cleaning line plug into the pressure transducer and place the line in aspiration valve.
  2. Dip the other end of the line in a cup of 50 cc of Isopropyl Alcohol or Hospital Grade Disinfectant.
  3. Press PRIME button and let Ocusystem run through entire priming cycle. Note: ADVANTAGE will stop on diagnostic #4. Press PRIME button again to continue.
  4. Refill the cup with 50 cc of the Isopropyl Alcohol and run another PRIME cycle.
Disinfect Pressure Transducer daily as described above:
Surgical Design Pressure Transducer

Back to top